Agilent 1200 G1312A Binary Pump
- Make Agilent
- Hertz 50Hz/60Hz
- Model G1312A
- Serial DE23912045
- Weight 18
- Voltage 230V
- Working Working
- Warranty AS-IS
- Dimensions 50×43×35 in
- Shipping Type Freight
- Item Condition Pre-owned
- Manufacturing Date N/A
Agilent 1200 G1312A Binary Pump
***The device turns on.That's all we can check.*
This is a refurbished Agilent 1200 Series G1312A Binary Pump.
The Agilent 1200 G1312A Binary Pump is a high-performance, binary pumping system for use in liquid chromatography. It features a compact design and operates using a binary pumping mechanism, which allows for high pressure operation and efficient sample delivery. The pump features a high-resolution, low-pulsation flow rate, making it ideal for use in demanding applications.
Type | Specification |
Hydraulic system | Two dual piston in series pumps with proprietary servo-controlled variable stroke drive, floating piston design and active inlet valve |
Setable flow range | Set points 0.001 – 5 ml/min, in 0.001 ml/min increments |
Flow range | 0.1 – 5.0 ml/min |
Flow precision | <0.07% RSD, or <0.02 min SD whatever is greater, based on retention time at constant room temperature |
Flow accuracy | +1% of 10µl/min whatever is greater |
Pressure | Operating range 0 400 bar (0-5880 psi) up to 5 ml/min |
Pressure pulsation | < 2% amplitude (typically < 1%), at 1 ml/min isopropanol, at all pressures > 1 MPa |
Compressibility compensation | User-selectable, based on mobile phase compressibility |
Recommended pH range | 1.0 – 12.5, solvents with pH <2.3 should not contain acids which attack stainless steel |
Gradient formation | High-pressure binary mixing, delay volume 180-480 µl without mixer, 600 – 900 µl with mixer, dependent on back pressure |
Composition range | 1 – 99 % or 5 µl/min per channel, whatever is greater |
Composition precision | <0.5% absolute |
Composition accuracy | 0.15% RSD, at 1 ml/min |
Control and data evaluation | Agilent Chem Station for LC |
Analog output | For pressure monitoring, 2mV/bar, one output |
Communications | Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional |
Safety and maintenance | Extensive diagnostics, error detection and display (through handheld controllers G4208 A, G1323 B and Agilent Chem Station), leak detection, safe leak handling, leak output signal for shut down of pumping system. Low voltage in major maintenance areas |
GLP features | Early maintenance feedback (EMF) for continuous tracking of instruments usage in terms of seal wear and volume of pumped mobile phase with user-settable limits and feedback messages. Electronic records of maintenance and errors. |
Testimonials
“REUZEit has been a great partner for our used equipment needs and always provide timely updates of new arrivals of consigned equipment.”
“Great company to work with. Tammy completed our first international equipment shipment with ease.”
“Fast response, open to adjusting schedule as needed, and great customer interaction.”