Thermo Dionex UltiMate ACC-3000T Thermostatted Autosampler Column CLEARANCE!
- Make Thermo Scientific
- Hertz 50Hz/60Hz
- Model Dionex Ultimate 3000 autosampler column compartment
- Serial 8161355
- Weight 25
- Voltage 120V/220V
- Working Working
- Includes Power Cord
- Warranty 30-Day Warranty, 100% Parts and Labor
- Dimensions 68×59×64 in
- Shipping Type Freight
- Item Condition Pre-owned
- Manufacturing Date 2019
Thermo Dionex UltiMate™ ACC-3000T Thermostatted Autosampler Column Compartment
Reduce the size and cost of your liquid chromatography (LC) system. The Thermo Scientific™ Dionex™ UltiMate™ ACC-3000 Autosampler Column Compartments provides a cost-effective solution for conventional LC and UHPLC applications at up to 62MPa (9000 psi) operating pressure. It integrates an autosampler and column compartment (with optional sample thermostatting) in a single module and offers performance and flexibility that exceed most analytical method requirements. The number of wear parts have been minimized for trouble-free operation.
Recommended Use
The UltiMate ACC-3000 Autosampler Column Compartments is recommended for cost-sensitive research and analytical laboratories. It is available with sample thermostatting (ACC-3000T) and without (ACC-3000) and is an essential part of the UltiMate 3000 Basic Automated System.
Performance
- Pulled-loop injection principle for full- and partial-loop injections.
- Wide injection volume ranges from 1 to 10µL in partial-loop and 20µL in full-loop mode, or 10 to 100µL and 200µL, respectively.
- Low sample carryover (typically < 0.01%).
- Built-in eluent preheating for improved chromatographic reliability at high temperatures.
- Patented, liquid-based temperature control for efficient heat extraction, protecting protect thermally sensitive samples.
- Accurate and precise column temperature control (±0.1°C measured in the sample) to meet qualification standards within pharma laboratories.
Ease of Use
- The Thermo Scientific™ Dionex™ Viper™ Fingertight Fitting system allows easy switching of injection volumes and sample loops.
- All fluidic components are accessible from the front.
For Use With (Equipment) | UltiMate 3000 Entry Level (Basic) |
Type | Thermostattted Autosampler Column Compartment |
Injection Cycle Time | <45 sec. |
Injection Volume Precision | <0.25% |
Injection Volume Range | 1 to 200 µL |
Injection Type | Pulled Loop |
Sample Capacity | 3 Racks |
Max. Pressure | 9000 psi (620 bar) |
Biocompatible Version | No |
Linearity | >0.9995 |
Voltage | 100 to 120 V |
Weight (Metric) | 23 kg |
Description | ACC-3000 T Thermostattted Autosampler Column Compartment |
Carryover | <0.02% |
Unit Size | Each |
Testimonials
“REUZEit has been a great partner for our used equipment needs and always provide timely updates of new arrivals of consigned equipment.”
“Great company to work with. Tammy completed our first international equipment shipment with ease.”
“Fast response, open to adjusting schedule as needed, and great customer interaction.”